Any establishment involved in the manufacture, preparation or distribution of a device can register with FDA. Registration itself is administrative. The reason it matters is that registration makes the site part of the regulated chain, which means the operations behind it must hold up to inspection: receiving, storage, handling, shipping and records.

Registration versus certification

FDA registration is an obligation, not an endorsement. ISO 13485 is a voluntary certification of the quality management system, and in practice it is the clearest evidence that a facility operates under documented, audited procedures. A warehouse can be registered without being certified, but the reverse is rarely true in a device supply chain.

What an investigator looks for

  • Documented procedures for receiving, storage, handling, shipping and returns.
  • Training records for the people performing those procedures.
  • Traceability records linking lot or serial number to location, transaction and destination.
  • Environmental controls and their monitoring records where storage conditions matter.
  • Quarantine and disposition records for damaged, expired or nonconforming product.
  • Complaint handling and correction records for devices the site distributed.

Part 820 in daily operations

The Quality System Regulation is often read as a manufacturing requirement. In a warehouse it shows up as control of nonconforming product, preservation of product, control of records, and corrective and preventive action. A site that can produce a signed deviation record and a completed CAPA is demonstrating more compliance than one that can only produce a certificate.

Questions worth asking a facility

Ask to see a sample traceability report, the temperature monitoring alerts, and a recent internal audit summary. The answers reveal more than the certificate on the wall: specifically, whether the records are produced routinely or reconstructed on request.